b'Performance you can trust.PRECISIONAccurate discrimination of fold changes is an important parameter for qPCR performance. To demonstrate this characteristic with X9, a 1.5-fold dilution series of synthetic cDNA was prepared, and 3 replicates of each dilution point run against 24 assays on a 192.24 Dynamic Array IFC. Amplification curves for the assay targeting GAPDH demonstrate accurate and precise quantitation of target abundance. High-precision performance. A 16-point dilution series of synthetic cDNA analyzed against an assay targeting GAPDH. Inset shows a magnified region of the amplification curves. REPRODUCIBILITYUniform amplification across replicates is a requirement for accurate quantitation by qPCR. To highlight the uniformity of amplification across an IFC, 96 replicates of a synthetic cDNA sample were run on X9 against96 assays using the 96.96 Dynamic Array IFC. A Ct standard deviation of 0.06 across 96 sample replicates on the IFC demonstrates excellent uniformity of amplification that will permit accurate quantitation of all samples, regardless of location on the IFC. Reproducible amplification. Amplification curves of 96 replicates of a synthetic cDNA sample run against an assay targeting GAPDH. Mean Ct is 12.4 with a standard deviation of 0.06 Ct. 13'